Georgieva, I. and Hofreither, C. and Ilieva, T. and Ivanov, T. and Nakov, S. (2013) Laboratory calibration of a MEMS accelerometer sensor. [Study Group Report]
|
PDF
1MB |
Abstract
In the present work we address MEMS accelerometers. There are many sources of error that affect their behaviour. In this report, we present two mathematical models and associated algorithms for calibrating the device, i.e. compensating for the constant bias error (the offset of its output signal, when the device does not undergo any acceleration), scaling errors (the deviation of the accelerometer’s scale from the unit we want to use, like m/s2 ) and nonorthogonality of the axis.
Item Type: | Study Group Report |
---|---|
Problem Sectors: | Technology/Devices Data processing |
Study Groups: | European Study Group with Industry > ESGI 95 (Sofia, Bugaria, Sept 23-27, 2013) |
Company Name: | MM Solutions AD |
ID Code: | 633 |
Deposited By: | Matthew Hennessy |
Deposited On: | 04 Dec 2013 23:28 |
Last Modified: | 29 May 2015 20:15 |